Product List

ALL PRODUCTS (A~Z)

Inquiry Now

  Email:

  hzck@com.hz.cn

  sales@cktrade.com.cn

 

 

   Fax:

  +86-571-87977703

 

 

  Tel:

  +86-571-87977725

  +86-571-87977729

  +86 18969038315

 

 

   Skype:

  ckskyphone

 

 

 

You are here: Home > Products > General > Testing Chamber > LHH SD/GSD series/SDP ser...Products

LHH SD/GSD series/SDP series/GP series/GSP series

Medicine Stability Testing Chamber

Microprocessor controller(with timing function)

Features:

1. Microprocessor control, stainless steel chamber, semicircular arcs at corners for easy cleaning

2. Even air circulating system

3. R134a refrigerant, 2 imported compressors and fan motor

4. Over temperature and temperature difference alarms

5. Imported humidity sensor which can be used in high humidity environment

6.  Balance temperature and humidity adjusting system

7.  There is a 25mm instruction connection hole on the left side of the chamber for easy testing operation and temperature measurement.

8. UV light system for periodic sterilization of chamber (Option)

9.  Independent audible and visible temperature-limiting alarm system ensures experiments run safely (Option)

10. RS485 connector can connect computer record and inspect the parameters and the variations of temperature (option)

 

Programmable Touch Screen

1.  Large LCD screen to display more data at same time

2.  English operation menu, display current data curves

3. 100 groups with 1000 period’s 999 circulations, max timing for each period is 99hours 59 minutes.

4. Auto lock after setting data.

5. Available to program on computer via RS-484 or RS-232

 

Safety device:

1. Compressor over-heat protection

2. Fan over-heat protection

3. Over-temperature alarm system

4. Over-compressing protection

5. Over-load protection

6.  Water lack protection

7.With temperature & Humidity

Model: LHH-150SDP/ LHH-250SDP/ LHH-500SDP

8.With temperature , Humidity & Light

Model: LHH-150GSP/LHH-250GSP, LHH-SSG

9.With temperature & Light

Model: LHH-150GP/LHH-250GP

 

Storage conditions for long-term retention sample stability test

Temperature: +25℃ ±2℃

Humidity: 60±5%RH

Time: 12 months

 

Storage conditions for accelerated stability test

Temperature: +40℃ ±2℃

Humidity: 70±5%RH

Time: 6 months

Illumination under strong light conditions: 4500±500LX

Above related parameter is for reference only.

 

Specifications: