You are here: Home > Products > Biological > Others > ESR-2068A Products

ESR-2068A

ESR Analyzer

Specification: 

Number of test samples: 10 items at the same time 
Max test speed: 20 samples/hour; 
Suited ESR tube: 1.6ml ESR tube(vacuum or normal); 
Sample volume: 1.6ml anticoagulant blood(blood 1.28ml + anticoagulant0.32ml),the height of the inputting tube 60mm; 
Accuracy range of sample: 50mm –64mm; 
Read resolution: ±0.2mm(westergren method result ±1mm); 
Interval of reading: 1 minute; 
Testing time: The test time is 30 minutes, equivalents to 1 hour westergren method; The test time is 60 minutes, equivalents to 2 hour westergren method; 
Display: 128x240 LCD screen; 
Power: 12V(SELV) by external power supply; 
Dimension: 310 x 190 x 180 (mm); 
Weight:4.8kg; 
 

Technical Parameter: 

Measure range:0—140 mm/h(westergren method); 
Testing pertinence:for westergren method result, the linear coefficient r≥0.98; 
Testing repeatability: CV < 3% ; 
Testing accuracy of crate temperature:no more than ±2.5°C between 15°C-30°C 
 

Feature:

Method of inputting sample: random sample position, testing at any moment ; 
Result Output:printing No, westergren method result and full ESR curve,testing time and temperature; if the testing time is 60 minutes, Kats value can be calculated, displayed, printed automatically; 
The device has the function to choose environment temperature compensating. If this function is valid, the test result under room temperature can be calibrated to 18°C by ESR calibrator.; 
The device can save more than 150 result , including sample No., ESR value and dynamic curve; 
The device has protection function when the power shuts down. The result will be saved automatically if the power shut down abruptly; 
The device is equipped with thermal printer, can work on manual or auto mode; 
Operation interface: panel keyboard; 
Communication interface:RS-232 serial port